Resumen:
he development of a sandwich-type biosensor for glucose quantification is presented. This work is focused on the optimization of the enzymatic matrix of the biosensor. The best performance was found for an enzymatic matrix composed by 30 % w/w mucin, 70 % w/w albumin, 1.35 U glucose oxidase (GOX) per sensor, and glutaraldehyde diluted to 3%. The crosslinking with glutaraldehyde transforms this mixture into a hydrogel that is entrapped between two membranes of polycarbonate. The selected sandwich-type biosensor showed very good response time, sensitivity, stability, and sensor-to-sensor reproducibility. According to the results presented in this manuscript, a biosensor prepared with very high amount of enzyme would not necessarily increase the analytical signal. Simulated curves are compared with experimental data to explain the dependence of sensitivity on the concentration of enzyme. In addition, this kind of comparison represents a quite simple way to estimate the value of vmax