GARAY FERNANDO SEBASTIÁN
Artículos
Título:
Mucin and carbon nanotube-based biosensor for detection of glucose in human plasma
Autor/es:
FAUSTO COMBA; MARCELO R. ROMERO; FERNANDO GARAY; ANA M. BARUZZI
Revista:
ANALYTICAL BIOCHEMISTRY
Editorial:
ACADEMIC PRESS INC ELSEVIER SCIENCE
Referencias:
Lugar: Amsterdam; Año: 2018 vol. 550 p. 34 - 34
ISSN:
0003-2697
Resumen:
his work reports an amperometric enzyme-electrode prepared with glucose oxidase, which have been immobilized by a cross-linking step with glutaraldehyde in a mixture containing albumin and a novel carbon nanotubes-mucin composite (CNT-muc). The obtained hydrogel matrix was trapped between two polycarbonate membranes and then fixed at the surface of a Pt working electrode. The developed biosensor was optimized by evaluating different compositions and the analytical properties of an enzymatic matrix with CNT-muc. Then, the performance of the resulting enzymatic matrix was evaluated for direct glucose quantification in human blood plasma.The novel CNT-muc composite provided a sensitivity of 0.44 ± 0.01 mA M−1 and a response time of 28 ± 2 s. These values were respectively 20% higher and 40% shorter than those obtained with a sandwich-type biosensor prepared without CNT. Additionally, CNT-muc based biosensor exhibited more than 3 orders of magnitude of linear dynamic calibration ran